| Location |
IMOS Capricorn Channel Mooring. Deployed 16/03/2011 |
| GPS |
22° 24.494' S, 151° 59.606' E |
| Deployment Type |
Mooring |
| Instrument |
WQM_121 |
| Description |
WETLABS, WQM, serial no. 121 |
| Capability |
OCEAN
FL 0-30
NTU 0-10
PRESS 0-350
Housing - 200m
|
| Sample Type |
CTD, DO, FLNTU |
| Sampling |
BLIS pump upgrade: N/A
Vacuum Tested: Y
TBT Replaced: Y
Flush & Clean Service: Y
Firmware: V1.23
DO stabilised 60s: Y
Pressure cutoff -99999.9: Y
1min/15min Samples
BLIS ON
Purge When Insitu On
BLIS Volume: 28µl
BLIS interval : 1hr
Free Disk: 1049 days
Memory Free: 1022767 Bytes
Battery: BPA50B #024
Bat Voltage 16.02
Setup file: 1211103.nsf
|
| Sample Depth |
17 (m from bottom) |
| Time : Switch On |
15/03/2011 at 20:58:20 UTC/GMT |
| Time : First Wet |
16/03/2011 at 00:51:00 UTC/GMT |
| Time : First Good |
16/03/2011 at 00:59:00 UTC/GMT |
| Time : Last Good |
12/08/2011 at 02:40:00 UTC/GMT |
| Time : On Deck |
12/08/2011 at 02:58:00 UTC/GMT |
| Comment |
Deployment: At startup nothing happened. Battery cable damaged - one pin is loose. Replaced cable with another which has a
cut in the sheath - repaired with amalgamating tape.
Recovery: No comms on recovery. Recovered data from card. DO sensor failed after 2 months. Other sensors worked for full deployment.
CTD due for calibration. FLNTU becomes noisy and then negative for last 20 days of deployment.
|
| Filename |
DATA/TRIP005139/WQM121_023.dat |